Power Supply - Swann AMI Hydrogen QED Operator's Manual

Table of Contents

Advertisement

AMI Hydrogen QED
Installation
3.6.
Installation
requirements
20

Power Supply

WARNING
Electrical shock hazard
Installation and maintenance of electrical parts must be per-
formed by professionals.
 Always turn off power before manipulating electric parts.
A
B
C
D
A
Power supply connector
B
Neutral conductor, Terminal 2
C
Phase conductor, Terminal 1
D
Protective earth PE
NOTICE: The protective earth wire (ground) has to be
connected to the grounding terminal.
The installation must meet the following requirements.
 Mains fuse 1.6 AT
 Mains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
 Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMI Hydrogen QED
A-96.250.791 / 040219