Swann AMI Turbitrace Operator's Manual page 46

Table of Contents

Advertisement

AMI Turbitrace
Maintenance
Calibration
flow
Manual
calibration
Automatic
calibration
44
LINE
A
A calibration can be started manually or automatically.To start a
manual calibration proceed as follows:
Maintenance
Verification
Calibration
Simulation
Set Time 01.01.05 16:30:00
Calibration
Manual operation
Parameters
Calibration
Method
Mode
fine
Method
Run Time
coarse
300 sec
Start Time
00:30:00
To program an automatic calibration proceed as follows:
Maintenance
Verification
Calibration
Simulation
Set Time 01.01.05 16:30:00
Cal
Empty
Zero
Line
CAL
EMPTY
1 Navigate to menu <Maintenance>/
3.2
<Calibration>.
2 Press [Enter].
3 Select <Parameters>.
4 Press [Enter].
5 Select <Method>.
3.2.2
 If measuring turbidity in pure
and ultrapure water, set method
to <fine>. For measuring
turbidity >0.5 FNU set method
to <coarse>.
6 Press [Enter], then press [Exit
7 Select <Yes> to confirm.
3.2.2.2
daily
8 Press [Enter]
fine
9 Select <Manual operation>
10 Press [Enter]
 The calibration starts.
1 Navigate to menu <Maintenance>/
3.2
<Calibration>.
2 Press [Enter].
B
A-96.250.131 / 050220