Instrument Specification - Swann AMI Turbitrace Operator's Manual

Table of Contents

Advertisement

AMI Turbitrace
Product Description
2.1.
Power Supply
Electronics
housing
Sample
requirements
On-site
requirements
10

Instrument Specification

Voltage:
Power consumption:
Aluminium with a protection degree of IP 66 / NEMA 4X
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
Flow rate:
Temperature:
Inlet pressure
Outlet pressure:
CAUTION
The measuring chamber is under the pressure of the
sample line
If the inlet pressure is too high, moisture can enter the housing
of the measuring electronics and lead to incorrect measure-
ments and failure of the instrument.
 The maximum inlet pressure of 10 bar must never be
exceeded, not even for short time periods. Take appropriate
measures to protect the instrument from pressure surges in
the sample line.
The analyzer site must permit connections to:
Sample inlet:
Sample outlet:
Max. Altitude:
100–240 VAC (±10%)
50 /60 Hz (±5%)
or 24 VDC (±10%)
max. 30 VA
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10 –90% rel., non condensing
backlit LCD, 75x45 mm
5–20 l/h
1–40 °C
1–10 bar
pressure free
Serto 4 x6 mm
Hose nozzle, 15 x20 mm
2000 m above sea level
A-96.250.131 / 050220