Product Description
The AMI Turbitrace is used for the measurement of low level turbid-
ity in pure and ultra pure water.
Turbidity is a measure of how much of the light traveling through
water is scattered by suspended particles. The scattering of light in-
creases with increasing suspended solid. The intensity of light scat-
tered at 90° as a beam of light passes through a water sample is
measured and indicates turbidity of the sample.
Two signal outputs programmable for measured values (freely scal-
able, linear or bilinear) or as continuous control output (control pa-
rameters programmable).
Current loop:
Maximal burden:
Third signal output available as an option. The third signal output
can be operated as a current source or as a current sink (selectable
via switch).
Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function. Both contacts can be used as normally open
or normally closed.
Maximum load: 1 A / 250 VAC
One potential free contact.
Alternatively:
 Open during normal operation, closed on error and loss of
power.
 Closed during normal operation, open on error and loss of
power.
Summary alarm indication for programmable alarm values and in-
strument faults.
For potential-free contact to freeze the measuring value or to inter-
rupt control in automated installations (hold function or remote-off).
No data loss after power failure. All data is saved in non-volatile
memory. Over voltage protection of in- and outputs.Galvanic sepa-
ration of measuring inputs and signal outputs.
 USB Interface for logger download.
 Third signal output (can be used in parallel to the USB interface)
 RS485 with Fieldbus protocol Modbus or Profibus DP
 HART interface
0/4–20 mA
510 Ω
7