AMI Turbitrace
Maintenance
40
1 Stop sample flow.
2 Turn the filter cover [A] out of the bayonet connection and re-
move it from the measuring chamber.
3 Pull out the zero-point filter.
4 Apply a thin film of PTFE grease to sealing surface of filter shell
and to zero-point filter O-ring.
5 Insert the zero-point filter.
6 Turn the filter cover [A] into the bayonet connection.
7 Open the sample flow.
A-96.250.131 / 050220