AMI Turbitrace
Program List and Explanations
2.2.2
2.2.2.1
2.2.2.1
2.2.3
2.2.3.1
2.2.3.1.1
2.2.3.2
2.2.3.2.1
2.3 Sample
2.3.1
2.4 I/O State
2.4.1- 2.4.2
2.5 Interface
56
Miscellaneous
Case Temp.: Temperature in °C inside the housing of the AMI
transmitter.
Dew point: Calculated dew point in °C inside the housing of the tur-
bidimeter.
History
Cal. history: Only for diagnostic purpose. Review the values of the
last verifications. Max. 64 data records are memorized.
Number: Verification counter.
Date, Time: Date and time of the verification.
Zero:
Flow deviation: Shows the flow deviation in %.
Ver. history: Only for diagnostic purpose. Review the values of the
last verifications. Max. 64 data records are memorized.
Number: Verification counter.
Date, Time: Date and time of the verification.
Actual value: The measuring value of the verification.
Setpoint: The value printed on the label of the verikit.
Deviation: Shows the deviation from the setpoint in % or FNU.
Sample ID: Shows the identification assigned to a sample. This
identification is defined by the user to identify the location of the
sample.
Sample flow: Flow in l/h
Shows the actual status of all in- and outputs.
Alarm Relay:
Relay 1 and 2:
Input:
Signal Output 1 and 2: Actual current in mA
Signal Output 3:
Only available if optional interface is installed.
Shows the programmed communication settings.
Active or inactive
Active or inactive
Open or closed
Actual current in mA (if option is installed)
A-96.250.131 / 050220