AMI Turbitrace
Program List and Explanations
5.3.4.3
5.3.4.4
5.3.4.5
5.4 Miscellaneous
5.4.1
5.4.2
A-96.250.131 / 050220
Output/Control: (relay or signal output):
Continuous:
Controller continues normally.
Controller continues based on the last valid value.
Hold:
Controller is switched off.
Off:
Fault:
No message is issued in pending error list and the
No:
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes:
Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.
Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6'000 Sec
Language: Set the desired language.
Language
German
English
French
Spanish
Set defaults: Reset the instrument to factory default values in three
different ways:
Set defaults
no
Calibration
In parts
Completely
 Calibration: Sets calibration values back to default. All other
values are kept in memory.
 In parts: Communication parameters are kept in memory. All
other values are set back to default values.
 Completely: Sets back all values including communication
parameters.