AMI Turbitrace
Program List and Explanations
3.4 Replace Absorber
3.5 Set Time

4 Operation

4.1 Sensors
4.1.1
4.1.2
4.2 Relay Contacts
4.3 Logger
A-96.250.131 / 050220
Press the [Enter] key.
 The value is simulated by the relay/signal output.
Alarm Relay:
Relay 1 and 2:
Input:
Signal Output 1 and 2: Actual current in mA.
Signal Output 3:
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
If maintenance message E068 or error message E006 appears,
call this menu item and replace the humidity absorber. This will
make both messages disappear for 3 days. This time is needed for
the new humidity absorber to dry the housing.
Adjust date and time.
Filter time constant: Used to damp noisy signals. The higher the fil-
ter time constant, the slower the system reacts to changes of the
measured value. Range: 5–300 sec
Hold after cal: To allow the instrument to stabilize again after cali-
bration. During cal. and hold time the signal outputs are frozen,
alarms and limits are not active. Range: 0–6000 sec
See chapter 5 Installation.
The instrument is equipped with an internal logger. The logger data
can be copied to a PC with an USB stick if option USB interface is
installed.
The logger can save approx. 1500 data records. Records consists
of: Date, time, alarms, measured value, temperature, flow.
Active or inactive
Active or inactive
Open or closed.
Actual current in mA (if option is installed)
59