AMI Turbitrace
Troubleshooting
Error
E017
Control Timeout
E018
Turbi disconnected
E024
Input active
E026
IC LM75
E028
Signal output open
E030
EEProm Frontend
E031
Calibration Recout
E032
Wrong Frontend
E033
Power-on
E034
Power-down
E068
Replace absorber
(dew point inside the tur-
bidimeter housing is 3 °C
or higher)
A-96.250.131 / 050220
Description
Corrective action
– check control device or programming in
Installation, Relay contact, Relay 1 & 2
5.3.2 & 5.3.3, p. 65
– shut off power
– check wiring
– See If Fault Yes is programmed in
Menu
5.3.4, p. 68
– call service
– check wiring on signal outputs 1 and 2
– call service
– call service
– call service
– none, normal status
– none, normal status
– call the <Maintenance>/<Replace
Absorber> menu item.
– replace the humidity absorber, see
Replacing the Humidity Absorber, p. 37
49