Power Supply - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Installation
3.9.
26

Power Supply

A
Power supply connector
B
Neutral N, Terminal 2
C
Outer conductor L, Terminal 1
D
Protective earth PE
Note: The protective earth wire (Ground) has to be connected
to the grounding terminal.
ATTENTION
It is not allowed to power external loads out of the
AMI transmitter.
Instrument power supply:
Only a certified electrician shall connect the instrument to
the mains.
 Fuse 1 AT
 Mains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
 Mains equipped with an external switch or circuit-breaker
- near the instrument,
- easily accessible to the operator,
- marked as interrupter for AMI Phosphate-II
A
B
C
D
A-96.250.641 / 290312