AMI Turbitrace
Program List and Explanations
4.3.1
Interval
1 s
Time
25 min
4.3.2

5 Installation

5.1 Sensors
5.1.2
5.2 Signal Outputs
5.2.1 and 5.2.2
5.2.1.1
5.2.1.2
5.2.1.3
60
Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the login buffer is full,
the oldest data record is erased to make room for the newest one.
(circular buffer)
5 s
1 min
2 h
25 h
Clear logger: If confirmed with yes, all data is erased and a new
data series is started.
Sensor type: Display of the used sensor type Trace
Dimension: Choose the measurement unit (FNU or NTU)
Signal Output 1 and 2: Assign process value, the current loop
range and a function to each signal output.
NOTICE: The navigation in the menu <Signal Output 1> and
<Signal Output 2> is identical. For reason of simplicity only the
menu numbers of Signal Output 1 are used in the following.
Parameter: Assign one of the process values to the signal output.
Available values: Meas. value, Sample flow.
Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
 Linear, bilinear or logarithmic for process values.
See
As process values, p. 61
 Control upwards or control downwards for controllers.
See
As control output, p. 62
5 min
10 min
5 d
10 d
30 min
1 h
31 d
62 d
A-96.250.131 / 050220