Cleaning the Test Unit

NOTICE: To avoid scratches on the surface of the test unit do
not use scrubbing materials.
After each verification dry the test unit with a soft, clean tissue.

Calibration

The AMI Turbitrace is factory calibrated with formazine and does
not require a recalibration. With this calibration the zero-point is set,
to compensate the signal drift due to a film building on the measur-
ing cell window.
For this purpose, the solenoid valve [A] is switched over and the
sample is led through the zero point filter [B].
NOTICE: When measuring turbidity in pure and ultrapure water
it is important to start a zero-point calibration periodically. This
has very little influence on the slope.