AMI Turbitrace
Maintenance
Replacement
procedure
38
A
A
Screw cover
B
New humidity absorber flask
C
Saturated humidity absorber
To replace the humidity absorber proceed as follows:
1 Select <Maintenance>/<Replace Absorber> and confirm with
[Enter].
 This makes the two messages E068 and E006 disappear for
three days.
2 Unscrew and remove the screw cover [A] from the measuring
cell [E].
3 Pull the flask [C] containing the saturated humidity absorber out
of the screw cover.
4 Remove the cover of the new humidity absorber flask [B].
5 Push the new humidity absorber flask into the screw cover.
B
C
D
Bayonet lock cover with
o-ring
E
Turbidimeter housing
D
E
A-96.250.131 / 050220