AMI Turbitrace
Maintenance
A-96.250.131 / 050220
Calibration
Manual operation
Parameters
Calibration
3.2.2.2
Mode
Mode
daily
interval
Method
daily
Run Time
300 sec
weekly
Start Time
00:30:00
off
Calibration
3.2.2.2
Method
Mode
daily
fine
Method
coarse
Run Time
300 sec
Start Time
00:30:00
Calibration
3.2.2.2
Mode
daily
Method
Run Time
300 sec
Start Time
00:30:00
3 Select <Parameters>.
3.2.2
4 Press [Enter].
5 Select <Mode> and press [Enter].
6 Select the <Mode> according your
requirements.
fine
 Additionally, the Modes <Input>
and <Fieldbus> are available
which you can chose by
scrolling down with the [
key.
7 Select <Yes> to confirm.
8 Select <Method>.
 If measuring turbidity in pure
and ultrapure water, set method
fine
to <fine>. For measuring
turbidity > 0.5 FNU set method
to <coarse>.
9 Select <Run time> and set it ac-
cording your requirements.
10 Depending on your settings set the
fine
interval or the start time.
]
45