Electrical Connections - Swann AMI Turbitrace Operator's Manual

Table of Contents

Advertisement

AMI Turbitrace
Installation
3.4.
Cable
thicknesses
Wire
A-96.250.131 / 050220

Electrical Connections

WARNING
Electrical hazard.
 Always turn off AC power before manipulating electric parts.
 Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
 Make sure the power specification of the instrument corre-
sponds to the power on site.
In order to comply with IP66, use the following cable thicknesses
A
A
PG 11 cable gland: cable Ø
B
PG 7 cable gland: cable Ø
C
PG 9 cable gland: cable Ø
NOTICE: Protect unused cable glands
 For Power and Relays: Use max. 1.5 mm
stranded wire with end sleeves.
 For Signal Outputs and Input: Use 0.25 mm
stranded wire with end sleeves.
B
C
5–10 mm
outer
3–6.5 mm
outer
4–8 mm
outer
2
/ AWG 14
2
/ AWG 23
19