Power Supply; Sensor; Swan Flow Meter - Swann AMU-II Oxytrace Operator's Manual

Table of Contents

Advertisement

AMU-II Oxytrace
Installation
3.3.
Installation
requirements
3.4.
3.5.
18

Power Supply

CAUTION
Do not apply power to the transmitter until all electrical
connections have been made.
A
B
C
The installation must meet the following requirements:
 Mains cable according to standards IEC 60227 or IEC 60245;
flammability rating FV1
 Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMU-II Oxytrace

Sensor

Terminals: see
Connection diagram, p.
Sensor settings: see

Swan Flow Meter

Terminals: see
Connection diagram, p.
17.
Instrument Setup, p.
17.
A
Pluggable terminal
block
B
Phase/(+) conductor
C
Neutral/(-) conductor
25.
A-96.150.541 / 111223