Load Firmware: Firmware updates should be done by instructed
service personnel only.
Load Firmware
no
yes
Password: Select a password different from 0000 to prevent unau-
thorized access to the following menus:
Messages
Maintenance
Operation
Installation.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representative.
Sample ID: Identify the process value with any meaning full text,
such as KKS number.
Line Break Detection: Define if message E028 should be issued in
case of a line break on signal output 1 or 2.
Choose between <Yes> or <No>.
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
Protocol: Profibus
Device address:
ID-Nr.:
Local operation:
Protocol: Modbus RTU
Device address:
Baud Rate:
Parity:
Protocol: USB-Stick
Only visible if an USB interface is installed. No further settings are
possible.
Protocol: HART
Device address:
Range: 0–126
Range: Analyzer; Manufacturer; Multivariable
Range: Enabled, Disabled
Range: 0–126
Range: 1200–115 200 Baud
Range: none, even, odd
Range: 0–63