AMI Turbitrace
Program List and Explanations
As process
values
5.2.1.40
5.2.1.40.10
5.2.1.40.20
5.2.1.40.11
5.2.1.40.21
A-96.250.131 / 050220
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
[mA]
20
B
10
12
0 / 4
0.0
0.1
0.2
A
linear
B
bilinear
[mA]
20
10
12
0 / 4
2
4
6
1
10
100
0
1
2
X Measured value (logarithmic)
Scaling: Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
Parameter Meas. value
Range low: 0.000–250 FNU/NTU
Range high: 0.000–250 FNU /NTU
Parameter Sample Flow
Range low: 0.0–100 l/h
Range high: 0.0–100 l/h
A
0.3
0.4
0.5
X Measured value
X
1'000
10'000
3
4
X
61