AMI Turbitrace
Troubleshooting
Error
E001
Alarm high
E002
Alarm low
E005
Range
E006
Absorber exhausted
(dew point inside the tur-
bidimeter housing is 5 °C
or higher)
E009
Sample Flow high
E010
Sample Flow low
E013
Case Temp. high
E014
Case Temp. low
48
Description
Corrective action
– check process
– check programmed value
64
– check process
– check programmed value
64
– turbidity out of range
– disappears if measuring chamber is
filled
– call the <Maintenance>/<Replace
Absorber> menu item.
– replace the humidity absorber, see
Replacing the Humidity Absorber, p. 37
– if the error appears again after a few
days, proceed as follows:
make sure that a new, unused hu-
midity absorber has been installed.
call the <Replace Absorber> func-
tion again and wait for another three
days. During this time, observe the
dew point and make sure that it de-
creases.
if the dew point does not decrease,
call service.
– check sample flow
– check programmed value
64
– check inlet pressure
– re-adjust the sample flow
– clean instrument
– check programmed value
64
– check case/environment temperature
– check programmed value
– check case/environment temperature
– check programmed value
5.3.1.1.1, p.
5.3.1.1.25, p.
5.3.1.2.2, p.
5.3.1.2.31, p.
5.3.1.4, p. 64
5.3.1.5, p. 64