Swann AMI Turbitrace Operator's Manual page 50

Table of Contents

Advertisement

AMI Turbitrace
Troubleshooting
Error
E001
Alarm high
E002
Alarm low
E005
Range
E006
Absorber exhausted
(dew point inside the tur-
bidimeter housing is 5 °C
or higher)
E009
Sample Flow high
E010
Sample Flow low
E013
Case Temp. high
E014
Case Temp. low
48
Description
Corrective action
– check process
– check programmed value
64
– check process
– check programmed value
64
– turbidity out of range
– disappears if measuring chamber is
filled
– call the <Maintenance>/<Replace
Absorber> menu item.
– replace the humidity absorber, see
Replacing the Humidity Absorber, p. 37
– if the error appears again after a few
days, proceed as follows:
 make sure that a new, unused hu-
midity absorber has been installed.
 call the <Replace Absorber> func-
tion again and wait for another three
days. During this time, observe the
dew point and make sure that it de-
creases.
 if the dew point does not decrease,
call service.
– check sample flow
– check programmed value
64
– check inlet pressure
– re-adjust the sample flow
– clean instrument
– check programmed value
64
– check case/environment temperature
– check programmed value
– check case/environment temperature
– check programmed value
5.3.1.1.1, p.
5.3.1.1.25, p.
5.3.1.2.2, p.
5.3.1.2.31, p.
5.3.1.4, p. 64
5.3.1.5, p. 64
A-96.250.131 / 050220