Program List and Explanations

3 Maintenance

3.1 Verification
3.1.1
3.2 Calibration
3.2.1
3.2.2
3.2.2.1
3.2.2.1
3.2.2.2
3.2.2.3
3.2.2.40
3.2.2.1
3.2.2.2
3.2.2.3
A-96.250.131 / 050220
Press [Enter] to start the verification and follow the instructions on
the screen. Further details see
Manual operation
Start calibration manually at any time.
Parameters
Mode: The following modes are available:
Mode
interval
The interval can be set between 1 to 12 h.
daily
Set the time of day to start a calibration.
weekly
Set the day(s) and time of day to start a calibration.
off
Automatic calibration is switched off.
Input
A calibration can be started via input.
Fieldbus
A calibration can be started via fieldbus.
Mode Interval
Method: The following methods can be chosen
Method
fine
For pure and ultra pure water
coarse
If the turbdity is >0.5 FNU
Run time: Measuring time, set the run time with the [
[
] key.
Range: 30–1800 sec
Interval: Set the interval with the [
Range: 1–12 h
Mode daily, details see
Method: see Mode interval,
Run time: see Mode interval,
Verification, p.
] or [
5.3.2.24, p. 67
3.2.2.1, p. 57
3.2.2.1, p.
57.
41.
] or
] key.
57