Cleaning The Measuring Chamber - Swann AMI Turbitrace Operator's Manual

Table of Contents

Advertisement

AMI Turbitrace
Maintenance
6.2.
Cleaning
36

Cleaning the Measuring Chamber

A
B
C
Line
A
Bayonet lock cover
B
Measuring chamber
C
Emitter LED
To clean the measuring chamber proceed as follows:
1 Close the pressure regulator valve to stop the sample flow.
2 Open the drain valve [F] to empty the flow cell.
3 Turn the cover [A] out of the bayonet connection and remove it
from the measuring chamber [B].
4 Clean the measuring chamber using a clean, soft tissue.
5 Clean the optical elements [C], [D] and [E] with another soft,
clean tissue.
6 Close the measuring chamber with the cover.
7 Close the drain valve.
8 Open the pressure regulator valve.
D
Scattered light sensor
E
Transmitted light beam sensor
F
Drain valve
A-96.250.131 / 050220
D
E
F