AMI Turbitrace
Program List and Explanations
5.3.2.24
5.3.2.340
5.3.2.44
5.3.2.54
5.3.2.6
5.3.2.7
5.3.2.24
5.3.2.341
5.3.2.44
5.3.2.54
5.3.2.6
5.3.2.7
A-96.250.131 / 050220
Interval
Interval: The interval can be programmed within a range
of 1–1440 min.
Run Time: Enter the time the relay stays active.
Range: 5–32400 Sec.
Delay: during run time plus the delay time the signal and control
outputs are held in the operating mode programmed below.
Range: 0–6'000 Sec.
Signal Outputs: Select operating mode of the signal output:
Signal outputs continue to issue the measured value.
Cont.:
Signal outputs hold the last valid measured value.
Hold:
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Signal outputs are switched off (set to 0 or 4 mA).
Off:
Errors, except fatal errors, are not issued.
Output/Control: Select operating mode of the controller output:
Controller continues normally.
Cont.:
Controller continues based on the last valid value.
Hold:
Controller is switched off.
Off:
daily
The relay contact can be activated daily, at any time of a day.
Start time: to set the start time proceed as follows:
1 Press [Enter], to set the hours.
2 Set the hour with the [
3 Press [Enter], to set the minutes.
4 Set the minutes with the [
5 Press [Enter], to set the seconds.
6 Set the seconds with the [
Range: 00:00:00–23:59:59
Run Time: see Interval
Delay: see Interval
Signal Outputs: see Interval
Output/Control: see Interval
] or [
] keys.
] or [
] keys.
] or [
] keys.
67