Program List And Explanations; Messages; Diagnostics - Swann AMI Turbitrace Operator's Manual

Table of Contents

Advertisement

AMI Turbitrace

Program List and Explanations

9.

1 Messages

1.1 Pending Errors
1.2 Message List
1.3 Maintenance List

2 Diagnostics

2.1 Identification
2.1.1
2.1.2
2.1.3
2.1.4
2.1.5
2.2 Sensors
2.2.1
A-96.250.131 / 050220
Program List and Explanations
Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
Shows the error history: Error code, date and time of issue, and
status (active, acknowledged, cleared).
64 errors are memorized. Then the oldest error is cleared to save
the newest error (circular buffer).
Provides the list of necessary maintenance. Cleared maintenance
messages are moved to the Message list.
In diagnostics mode, the values can only be viewed, not modified.
Designation: of instrument: AMI Turbitrace
Version: firmware version, e.g. V6.20-09/16
Version TURBI:, e.g. 1.B4
Factory test: date of instrument, mother board and front-end QC
test.
Operating time: years, days, hours, minutes, seconds
Turbidity:
Turbidity: in FNUor NTU
(Raw value): Turbidity in counts.
(Quotient): Division of detector signal by reference signal.
Scale factor 1: correction factor 1 determined during factory
calibration.
Scale factor 2: correction factor 2 determined during factory
calibration.
55