Adjust The Sample Flow With A Capillary Tube - Swann AMI Turbitrace Operator's Manual

Table of Contents

Advertisement

AMI Turbitrace
Installation
3.3.3
A-96.250.131 / 050220

Adjust the Sample Flow with a Capillary Tube

The AMI Turbitrace is as standard delivered with the capillary FEP
tube with an inner diameter of 1 mm and a length of 500 mm.
If you have a low sample pressure or you need a high sample flow,
a shorter capillary with 186 mm length can be used.
A
B
C
D
A
Pressure regulator valve
B
Tube fitting
To replace the capillary proceed as follows:
1 Pull out the knob of the pressure regulator valve [A].
2 Close the valve.
3 Push in the knob to lock the valve in closed position.
4 Unscrew and remove the tube fittings [B] of the capillary [C].
5 Then screw in the tube fittings of the 186 mm capillary.
6 Tighten them well.
7 Open the pressure regulator valve.
C
500 mm capillary
D
186 mm capillary
17