The sample flows through the sample inlet [F] via solenoid valve
[G] and the manual 5 way valve [E] into the measuring chamber [B],
where the turbidity is measured. From there it flows through the
sample outlet [C] and the pressure regulator [K]. The pressure reg-
ulator is used to regulate the sample flow. Then the sample flows
via capillary [L] and flow meter [M] into the drain funnel [N].
To keep the measuring chamber clean and free from deposits it can
be cleaned by opening the drain valve [D].