Main Display - If VTBI has not been set (flow sensor must be used):
Infusion Status/
Drug Name/Profile
Name/Primary or
Secondary
(Only if secondary
is enabled in the data set)
Infusion Rate
Volume Infused
Softkey Identifiers
Softkeys
Main Display - If VTBI is set:
Infusion Status/
Drug Name/Profile
Name/Primary or
Secondary
(Only if secondary
is enabled in the data set)
Infusion Rate
Dose Rate
Volume to be Infused
Volume Infused
Time Remaining
Softkey Identifiers
Softkeys
Screen icon:
w
?
!
Main Display Features
ON HOLD
RATE
25.0
VOLUME
VOLUME
Clear Volume
Infused
Set VTBI Option
ADRENALINE
RATE
25
.0
16.
7
VTBI
VOLUME
1h
48m
VOLUME
Clear Volume
Infused
Set VTBI Option
TIME REMAINING DISPLAY icon - Indicates time remaining before VTBI will be completed. If
the time is greater than 24 hours then 24+ will be displayed.
PRESSURE INFORMATION icon - Shows the pressure from level 0 being the first bar to level 8.
Alarm limits: level 2, 5 or 8.
Indicates that a Guardrails® safety protocol is not in use. CareFusion recommends the use of
Guardrails® safety limits when setting infusions as standard practice.
Indicates that the value entered is outside of the soft limits. The warning can be overridden.
(Indicates Guardrails® safety protocol is in use)
Indicates that the value entered is outside of the hard limits. The warning can NOT be
overridden. This symbol is also used to prompt the user to set the rate. (Indicates Guardrails®
safety protocol is in use)
1000DF00498 Issue 3
w
ml/h
50
.0
ml
VTBI
ml/h
µg/kg/24h
45
.0
ml
50
.0
ml
00s
VTBI
7/42
ON HOLD
a)
SET RATE WITH
RATE
If the rate has not been set
ON HOLD
ON HOLD
and is showing at 0.0ml/h,
RATE TOO LOW
SET RATE WITH
then message a) will be
RATE
RATE
displayed.
ON HOLD
ON HOLD
RATE TOO LOW
RATE TOO HIGH
b)
ON HOLD
RATE
RATE
SET RATE WITH
RATE
If programmed rate is
ON HOLD
between 0.0ml/h and
RATE TOO HIGH
ON HOLD
0.1ml/h exclusive in drug
RATE
protocol, message b) will
RATE TOO LOW
be shown.
RATE
ON HOLD
RATE TOO HIGH
c)
RATE
If programmed rate is
greater than the Infusion
Rate Max in drug protocol,
message c) will be shown.