Two point oxygen calibration

1. Adjust flow meters and turn on driver.
2. Press the Tool button in the Setup screen.
3. Adjust the Oxygen Control to 21%:
a. Press the flashing button to confirm.
b. Allow reading to stabilize.
c. A check mark appears and %O
read 21%.
4. Adjust the Oxygen Control to 100%:
a. Press the flashing button to confirm.
b. Allow reading to stabilize.
c. A check mark appears and %O
read 100%.
5. Press Exit button.
display should
2
display should
2
7