Using Automatic Calibration - GE Moisture Image 1 Series Programming Manual

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Advertisement

Using Automatic
Calibration
Advanced Programming
The Series 1 automatically calibrates the moisture, pressure, and
oxygen measurement circuitry (Auto-Cal) at a user selected interval.
Auto-Cal compensates for any drift in the electronics. GE
Panametrics recommends setting the Auto-Cal interval to eight hours.
Set a smaller interval to Auto-Cal more frequently if the meter is
exposed to extreme temperatures or weather conditions.
Note: If you are using only Moisture Image Series Probes to
measure moisture, you do not have to Auto-Cal.
To set the Auto-Cal interval, press the keys shown in Table 3-3. The
screen appears similar to Figure 3-3 below. Refer to Figure 3-19 on
page 3-32 for a menu map.
Table 3-3: Entering the Reference Menu
Press the Main Menu key:
SETTINGS*
SERVICE
REFS
* The Settings Menu will prompt for a passcode (see page 2-6).
MOISTURE REFERENCE TABLE
Ch
HIGH
1
2
3
4
5
6
AutoCal Interval: 96 minutes.
Reference Menu
HELP
INTERVAL
Figure 3-3: Setting the Auto-Cal Interval
April 2004
To enter the:
Settings Menu
System Service Menu
Reference Menu
LOW
PROBE
DONE
3-7