Entering Auxiliary Input
Calibration Data
Basic Programming
Note: If you are using an auxiliary input to measure pressure, enter
pressure calibration data as described in Entering Pressure
Calibration Data as described on page 2-17.
This section describes how to enter calibration data for an auxiliary
input (optional). If your device is not supplied by GE Panametrics,
you will have to obtain the calibration data in order to properly set up
the Series 1.
You will need to enter two to 16 data points. If the sensor is linear,
you only need to enter two data points. Each data point requires a
voltage or current value depending on the probe configuration (see
page 2-9) and a corresponding scale value. Use the procedure below
to enter data.
Figure 3-19 on page 3-32 for a menu map.
IMPORTANT: Enter the auxiliary calibration data on the Program
Information List (Appendix A of the Startup Guide)
and store it in a safe place. If it is lost, contact GE
Panametrics for a duplicate.
1. At the System Calibration Menu (see Table 2-3 on page 2-12),
press the PROBE menu key until the Auxiliary Input Calibration
screen appears. There is one screen each for Auxiliary Input 1 and
Auxiliary Input 2. Make sure you have the correct screen before
verifying or entering data (see Figure 2-10 below).
Aux 1 Probe Calibration
ND:
Label:[No Label]
#
mA (VDC)
01
02
03
04
05
06
07
System Calibration Menu
HELP
CHANNEL
Figure 2-10: System Calibration Menu for Auxiliary Inputs
[ Aux # ]
##
mA
08
09
10
11
12
13
14
15
PROBE
April 2004
Ch
1
[ Aux #]
DONE
2-19