Entering Delta F Cell
Calibration Data (cont.)
2-18
Note: If you are using an auxiliary input to measure pressure, the
first column will be in mA or Volts depending on how you
configured the auxiliary input in Activating and Changing
Probes as described on page 2-9.
2. Use the CHANNEL menu key to cycle to the desired channel. The
channel number is indicated in the top right-hand corner of the
screen. The screen will only display installed channels.
S/N
3. Move the pointer to
4. Enter the probe serial number from the Calibration Data Sheet and
press [YES].
5. Move the pointer to the
6. Enter the voltage or current value and press [YES].
7. Move the pointer to the
8. Enter the PSIg value and press [YES].
9. Use Steps 5 through 8 to enter the span values.
Repeat steps 1 through 9 to enter pressure calibration data for the
remaining channels.
To enter calibration data for other probes, refer to Entering Moisture
Probe Calibration Data on page 2-13, Entering Delta F Oxygen Cell
Calibration Data on page 2-15, and Entering Auxiliary Input
Calibration Data on page 2-19. To exit, press the DONE menu key
until Main Menu appears on the message line. Then refer to Entering
High and Low Reference Values on page 2-21.
and press [YES].
Zero mV (mA)
line and press [YES].
Zero PSIg
line and press [YES].