Testing Recorder Outputs - GE Moisture Image 1 Series Programming Manual

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Advertisement

April 2004
Testing Recorder
Outputs
4-10
1. To test the recorder outputs, click on the [
"Recorder Properties" window (see Figure 4-5 on page 4-9). A
message appears: "This will suspend Recorder #[X] on Channel
[X]. Are you sure you want to test this recorder?" Click on [
continue.
2. The "Test Recorder" window appears, as shown in Figure 4-6
below. At the top, slide the pointer to the desired percentage of the
full scale recorder range, and click on [
3. To trim the recorder outputs, scroll the arrows to the right of the
"Zero" and "Span" windows up or down until you reach the
desired values, and click on [OK].
Note: For a complete discussion of testing and trimming recorder
outputs, refer to the sections Testing the Recorder Outputs
and Trimming Recorder Outputs in Chapter 2 of the Service
Manual.
The screen returns to the "Recorder Properties" window.
Figure 4-6: Recorder Test Window
] button in the
Test
OK
].
OK
Programming with PanaView
] to

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mis 1

Table of Contents