Entering High And Low Reference Values - GE Moisture Image 1 Series Programming Manual

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Advertisement

Entering High and Low
Reference Values
Basic Programming
The last step in setup is entering high and low reference values. The
Series 1 requires reference values for its measurement circuitry for
moisture, oxygen, and pressure inputs. The references are factory
calibration values that are specific to each channel card. Reference
values are located on a label placed on the side or front of the unit.
Note: If you receive a channel card replacement or change the
instrument program, you need to re-enter the reference data
for that channel.
Table 2-4 shows the key sequences for entering the Reference Menu.
(Figure 3-19 on page 3-32 for a menu map.)
Table 2-4: Moving to the Reference Menu
Press the Main Menu key:
SETTINGS*
SERVICE
REFS
*The Settings Menu will prompt for a passcode (see page 2-6).
Use the appropriate section that follows to verify or enter reference
values for the probes.
Once you verify and/or enter reference data for each input, refer to
Displaying Measurements on page 2-28.
IMPORTANT: You should record this data on the Program
Information List supplied in Appendix A of the
Startup Guide, or on a separate sheet of paper, and
keep it in a safe place.
April 2004