Setting Error Processing; Procedure For Setting Error Processing - GE Moisture Image 1 Series Programming Manual

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Advertisement

January 2004
Setting Error
Processing
Procedure for Setting
Error Processing
3-28
The Series 1 user program includes an Error Processing menu that
may be used to specify how the recorders/analog outputs and the
alarms respond to Range, Signal and Calibration errors.
The meter categorizes range errors as Under Range if the input
signal goes below the probe calibration range or Over Range if the
input signal goes above the probe calibration range. However, signal
errors are always categorized as Over Range, while calibration
errors are always categorized as Under Range.
Note: See Chapter 2 in the Service Manual for a detailed description
of Range, Signal and Calibration Errors.
When an error condition occurs, the meter displays an error message
on the screen and it drives the recorders/analog outputs and alarm
relays as indicated in Table 3-10 below.
Table 3-10: Error Indicators
Error Type
Range Error
Signal Error
Calibration Error
* "Mode" indicates one of the available measurement modes
** currently programmed error processing setting (see next section)
IMPORTANT: The factory default setting for Error Messaging is
, but the factory default settings for both Under
ON
Range and Over Range errors is
the recorder/analog outputs to 0 mA or 0V).
The Series 1 user program allows you to turn Error Messaging either
or
. In addition, if Error Messaging is
ON
OFF
separate Over Range and Under Range responses for all alarm relays
(if installed) and recorder/analog outputs. The error responses are
independently programmable for each channel.
Table 3-11 on the next page provides instructions for entering the
Error Processing Menu, and Figure 3-17 on the next page shows a
typical menu screen. To suppress or activate the Display Messages
function, move the pointer to the Display Messages line and use the
menu key to select
ACTION
If Display Messages is set to
up error processing:
Screen Error
Output & Alarm
Message
Relay Response
Over Range
Over Range Setting**
Under Range
Under Range Setting**
"Mode" Fault!*
Over Range Setting**
Cal Error
Under Range Setting**
ON
(the default) or
ON
OFF
ON
, use the steps on the next page to set
Advanced Programming
(this forces
Ignore
, you can program
.