Entering Pressure
Reference Values (cont.)
Basic Programming
HIGH
3. Move the pointer to
[YES].
4. Enter the high reference value for that channel and press [YES].
5. The pointer automatically moves to
Press [YES].
6. Enter the low reference value for that channel and press [YES].
7. Repeat steps 3 through 6 to enter the high and low reference
values for the remaining channels.
To enter reference values for moisture and oxygen inputs, refer to
Entering Moisture Reference Values on page 2-22, and Entering the
Delta F Oxygen Cell Reference Values on page 2-25. To exit, press
the DONE menu key until Main Menu appears on the message line.
Then refer to Displaying Measurements on page 2-28. When you exit,
the Series 1 automatically calibrates each channel.
for the desired channel and press
LOW
for the same channel.
April 2004
2-27