Specifying User
Functions
Selecting Initial
Parameters
Advanced Programming
User functions enable operators to program up to four desired
mathematical equations on each channel. They can also use any
parameter on any channel to calculate a different parameter. Use
Table 3-5 to enter this menu, and press the PAGE menu key until the
screen is similar to Figure 3-5 below.
Table 3-5: Entering the User Function Menu
Press the Main Menu key:
OPTIONS
USER
PAGE
SPECIFY USER FUNCTION
Function: User 1
Decimal Places: 1
Units Symbol: TEST
Label:
Equation:
log (DP°/C-15.3)
User Function Menu
HELP
CHANNEL
Figure 3-5: User Function Menu
1. Press the CHANNEL key to scroll to the desired channel, and the
PAGE key to scroll to the desired function (or blank page).
2. Move the pointer to the Function line. Press [YES] to remove the
function, type in a new entry, and press [YES] to confirm the
entry.
3. Move the pointer to the Decimal Places line. Press [YES] to erase
any current number, and enter the number of decimal places to the
right of the decimal point that you wish to appear. Press [YES] to
confirm the entry.
4. Move the pointer to the Units Symbol line. Press [YES] to remove
any entry, type in a new entry, and press [YES] to confirm the
entry.
5. Move the pointer to the Label line. Press [YES] to remove the
function. Type in a new entry, and press [YES] to confirm the
entry.
To enter the:
Option Menu
User Constant Menu
User Function Menu
Ch
TT
PAGE
April 2004
1
DONE
3-11