Entering Calibration
Data (cont.)
Entering Moisture Probe
Calibration Data
Basic Programming
Refer to the appropriate section that follows to verify or enter
calibration data for:
Moisture Probes, page 2-13
Delta F Oxygen Cell, page 2-15
Pressure Sensor or Transmitter, page 2-17
Optional Auxiliary Input(s), page 2-19
Be sure you have the Calibration Data Sheets that are supplied with
each GE Panametrics probe. Each Calibration Data Sheet consists of
a list of data points that you will need to enter or verify. Each
Calibration Data Sheet lists its corresponding probe serial number, as
well as the preassigned channel number. Calibration Data Sheets are
usually packed inside the probe cases.
Once you verify and/or enter calibration data for each input, refer to
Entering High and Low Reference Values on page 2-21.
IMPORTANT: Staple Calibration Data Sheets to the Program
Information List (Appendix A of the Startup Guide)
and store them in a safe place. If they are lost,
contact GE Panametrics for a duplicate. If you alter
any of the calibration data, make sure you make the
change on the Calibration Data Sheet.
You only need to enter calibration data for M and TF Series probes.
It is not necessary to enter calibration data for the Moisture Image
Series Probe unless you send the probe back to GE Panametrics for
calibration without its electronics module. If this is the case, enter the
calibration data for the recalibrated probe as described in this section.
The Series 1 will automatically download the new calibration data
into the Moisture Image Series Probe electronics module.
Note: You do not need to enter calibration data for the Moisture
Image Series Probe because it is stored in the probe's
electronics module. The Moisture Image Series Probe uploads
the calibration data into the Series 1 memory when needed.
Use the procedure below to enter the following data:
the probe serial number
the number of data points
the dew point and MH (or FH, depending on the probe) reading for
each data point