GE Moisture Image 1 Series Programming Manual page 106

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Advertisement

April 2004
Changing Alarm
Settings (cont.)
4-8
4. To set the alarm type, click on the [
radio buttons, as desired.
5. To determine the alarm values, enter the Trip and Deadband
values in the designated text boxes. (For a definition of the
Deadband value, see Setting Up the Alarms on page 3-4.)
6. To determine how the alarm will react when it encounters an error:
a. Pull down the "High" list box under the "On Error" line.
b. Then click on the desired option from the listing: "Ignore
error," "Trip on error," "Reset on error," or "Hold on last
error."
c. Repeat the procedure for the "Low" list box.
7. To test the alarm, click on the [
button to trip the alarm, and on the [Reset] button to reset it.
8. Click on [
] to immediately apply the changes, and on [
Apply
confirm the changes and close the window. (To return the window
to its previous settings, click on the [
] or [
Trip Above
Trip Below
] button. Click on the [
Test
] button.)
Cancel
Programming with PanaView
]
]
Trip
] to
OK