Entering Delta F Oxygen
Cell Reference Values
(cont.)
Basic Programming
3. Move the pointer to the
desired channel and press [YES].
4. Enter the zero value for the high reference for that channel and
press [YES]. The pointer automatically proceeds to the
for the high reference for the same channel. Press [YES] again.
5. Enter the span value for the high reference for that channel and
press [YES].
6. Move the pointer to the
desired channel and press [YES].
7. Enter the zero value for the low reference for that channel and
press [YES].
8. Move the pointer to the
desired channel and press [YES].
9. Enter the span value for the low reference for that channel and
press [YES].
Repeat steps 3 through 9 to enter the high and low reference values
for the remaining channels.
To enter reference values for moisture and pressure inputs, refer to
Entering Moisture Reference Values on page 2-22 and Entering
Pressure Reference Values on page 2-26. To exit, press the DONE
Main Menu
menu key until
to Displaying Measurements on page 2-28. When you exit, the
Series 1 automatically calibrates each channel.
Zero
line for the high reference for the
Zero
line for the low reference for the
Span
line for the low reference for the
appears on the message line. Then refer
April 2004
Span
line
2-25