5 Clinical Configuration and Use
5.1 Accessing the clinical screens
Warning
!
Only physicians and competent nursing and technical staff are permitted to access the
clinical screens. The patient must never be told how to access these screens.
Monday 18
Monday
P1: P1:
P2: P2:
Figure 5.1: Accessing the clinical screens.
To access the clinical screens, press the closed padlock icon
•
Pressing the icon for 5seconds (1 beep) will unlock access for a period of 4 minutes from the
last action on the screen (pressing the Alarm Silence button, the touch screen or the On/Off
button will reset the timer)
•
Pressing the icon for 8seconds (2 beeps) will unlock access for a period of 60 minutes from
the last action on the screen (pressing the Alarm Silence button, the touch screen or the
On/Off button will reset the timer). The time remaining is displayed on the padlock
•
Pressing the icon for 11seconds (3 beeps) will unlock access permanently. The infinity symbol
can be seen on the padlock
Caution
!
To relock the screen, press the open padlock icon or disconnect the ventilator from the
mains supply and switch it off.
NOT013189-6
18 June
June
9:58:46
9:58:46
Intern.
In
tern.
ACV
ACV
START
START
PS
PS
VENTIL
VENTIL
1s 1s
;
.
∞
Last test the 18/06/07
TEST
TEST
Double circuit : PASS
P1: P1:
P1: P1:
P2: P2:
PATIENT
PATIENT
:
Clinical Configuration and Use
ACV
ACV
PS
PS
NEW
NEW
;
60
63/174