ResMed Elisee 150 Clinical Manual page 117

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

Message
INT BATT
LOW.
CIRCUIT
OPEN.
CIRCUIT
OPEN.
TEST FAILED.
MAXI F .
MAXI Leaks.
WRONG CIRCUIT.
LOW PEEP .
MAXI P expi.
NOT013189-6
Icon
Alarm priority
Medium priority.
INT.
High priority.
LP LP
High priority.
LP LP
High priority.
Low priority.
F F
High priority.
High priority.
Medium priority.
LP LP
High priority.
HP
HP
Cause
Remedy
Internal battery
Change power
charge less than
supply.
15%.
Faulty connection
Reconnect patient
of patient circuit.
circuit.
Defective patient
Change patient
circuit.
circuit.
Manual test was
Repeat manual
interrupted.
test.
Respiratory rate
Check that
higher than upper
Respiratory Rate
threshold for
alarm and
Respiratory Rate
inspiratory trigger
alarm.
settings are
correct.
Leak higher than
Check that High
the High Leak
Leak alarm
alarm threshold.
threshold setting
is correct and
check patient
circuit integrity.
Type of patient
Check circuit and
circuit used is
repeat test.
different from the
one determined
by the test.
PEEP delivered to
Check patient
patient is lower
circuit integrity.
than target PEEP .
Expiratory
Check expiratory
pressure too high
valve control and
compared to
check patient
target PEEP .
circuit for
blockage.
Alarms
117/174