External Dc Power Supply - ResMed Elisee 150 Clinical Manual

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

External DC power supply

External power supply cord (mains
12-28 VDC power cord
power pack/Elisée connection)
Figure 2.8: External DC power cords.
An external DC power source (in a vehicule with 12-28V, 15A max.) can be used by connecting
the external power cord to the right side of the ventilator (see "Figure 3.3: 12-28VDC power cord,
connected." on page 24).
Note: The external DC power source must have a guaranteed life of at least one hour. If the
external power supply is less than 19 +/- 1V, the batteries will not be recharged.