Alarm
Minimum
FiO2.
Maximum
FiO2.
Leak.
Apnea.
Minimum
Expiratory
Pressure.
Maximum
Respiratory
Rate.
*In pressure mode, the Pmax alarm threshold cannot be set to a value lower than the delivered
Pinsp (or the delivered pressure support). If the user sets an alarm threshold lower than this
pressure value, the alarm threshold setting will be ignored. In this situation, the delivered Pinsp
(or pressure support) becomes the Pmax alarm threshold.
**When using a single circuit in a non-invasive mode, there is no Low PEEP alarm.
***When using a single circuit, the Pinsp (Pmax - PEEP) target must be set at a level at which
disconnection of the patient circuit will be detected.
****Adult: 15s below the threshold. Paediatric: 15s – (Measured respiratory rate/5), min. 5s
below the threshold.
NOT013189-6
Icon
Message
MINI FiO2.
O O
2
MAXI FiO2.
O O
2
MAXI Leaks.
PATIENT
APNEA.
LOW PEEP**.
LP LP
MAXI F .
F F
Δ
Alarm
Trigger
priority
threshold
High priority.
Alarm setting.
High priority.
Alarm setting.
High priority.
Alarm setting.
Medium
Tapnea.
priority.
Medium
PEEP - 1mbar.
priority.
Low priority.
Alarm setting.
Time to trigger
30 seconds
below alarm
threshold.
30 seconds
above alarm
threshold.
6 consecutive
cycles above
alarm
threshold.
Immediate.
30 cycles
below alarm
threshold.
4 consecutive
cycles
above alarm
threshold.
Alarms
115/174