Testing settings alarms for use with a double limb circuit
Connect a double limb patient circuit fitted with a MAQUET test lung to the ventilator and perform
the manual test.
Set up ACV mode, non-invasive, adult ventilation.
Set the ventilation parameters as follows:
•
V
T
•
PEEP = 0cmH
•
F = 15bpm;
•
Max flow ( ) = 23L/min (Ti = 1.3 or I/E = 1/2.0);
•
Trigger =
•
Flow shape = constant;
•
Start ventilation and perform the following operations.
Table 7.10: Testing settings alarms for use with a double limb circuit
Alarm
MAXI Leaks.
MINI Vt.
MAXI Vt.
124/174
= 500mL;
O;
2
V
;
NO
Action
Adjust alarm
upper threshold to
Leak = 20% then
create a slight
leak in the circuit
(e.g. by slightly
loosening the
expiratory limb of
the circuit).
Adjust alarm
upper threshold to
Leak = 80% and
stop the leak.
Press Alarm
Silence button.
Adjust alarm
upper threshold to
V
= 2000mL
TE
and lower
threshold to V
800mL.
Adjust alarm
lower threshold to
V
= 150mL.
TE
Press Alarm
Silence button.
Adjust alarm
upper threshold to
V
= 250mL.
TE
Adjust alarm
upper threshold to
V
= 2000mL.
TE
Press Alarm
Silence button.
Visual alarm
status
Flashing.
Steady.
Off.
Flashing.
=
TE
Steady.
Off.
Flashing.
Steady.
Off.
Audible alarm
Time taken for
status
alarm to trigger
Sounding.
6 cycles.
Off.
Off.
Sounding.
5 cycles.
Off.
Off.
Sounding.
5 cycles.
Off.
Off.
NOT013189-6