Pressurised Air Specifications; Specifications And Performance Data - ResMed Elisee 150 Clinical Manual

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

Pressurised air specifications

Low-pressure O
Maximum input pressure = 400kPa.
Maximum input flow = 15L/min.
Pressurised air interface
Female connection (built-in): 1.8 NPT threaded with clapper valve. CPC reference: MCD10-02.
Male connection (cord): CPC reference: MC22-03.

Specifications and performance data

Performance data
Ventilator weight: 3.7kg (9.9lbs).
Mains power pack weight: 0.45kg (0.88lbs).
External battery pack weight: 0.75kg (2.42lbs).
Dimensions: 290 x 250 x 130mm (11.41'' x 9.84'' x 5.11'').
Sound level: < 40dBa (at 1metre).
Mechanic and pressurised air performance: 0 – 60hPa (= 61.0cmH
of 0 - 180L/min.
Air pressure supply (minute volume) adjustment range: Adult: 0.20 – 30L/min.
Paediatric: 0.20 - 20L/min.
Monitoring device: black box (events data storage).
Ventilation modes
Volume-Controlled Ventilation (CV).
Pressure-Controlled Ventilation (PCV).
Assisted Volume-Controlled Ventilation ACV.
Pressure Assisted Controlled Ventilation PACV.
Intermittent Assisted Volume-Controlled Ventilation (Synchronised intermittent mandatory
ventilation, SIMV).
Intermittent Assisted Pressure-Controlled Ventilation (Pressure Synchronized Intermittent
Mandatory Ventilation, PSIMV).
Pressure Support PS.SV (PS).
Pressure Support with Minimum Tidal Volume PS.VT.
Display
Built-in touch screen.
Liquid Crystal Display (LCD).
142/174
port
2
O) at a flow rate
2
NOT013189-6

Advertisement

Table of Contents
loading

Table of Contents