Note: During (A)CV, the Elisée 150 delivers flow-controlled cycles. Airway pressure and alveolar
pressure then depend on the resistance and compliance of the patient's respiratory system.
It is important to set the high-pressure alarm (Pmax) to avoid the risk of barotrauma.
The bar graph, shown on all ventilation screens, allows you to monitor airway pressure.
CC
P
V
Fixed Ti
Figure 4.18: CV mode - Controlled Volume Ventilation.
CC
P
V
Ti fixed
Figure 4.19: ACV mode - Assisted Controlled Volume Ventilation.
NOT013189-6
CC
T mecha
T mecha
Fixed Ti
T mecha: Device period (depends on frequency)
CC: Controlled Cycle
ACC
T mecha
T mecha
Ti fixed
T mecha: Device period (depends on frequency)
ACC: Assisted Controlled Cycle
CC: Controlled Cycle
CC
T mecha
Fixed Ti
Fixed Ti
CC
CC
T mecha
Ti fixed
Ti fixed
CC
T mecha
T mecha
Ventilation
t
t
t
t
t
49/170