Table 7.5: Audible Alarms Which Can Be Turned Off - ResMed Elisee 150 Clinical Manual

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

Message
MAXI P inspi.
MAXI Vt.
MINI Vt.

Table 7.5: Audible alarms which can be turned off

Message
EXT BATT
MISSING.
EXT MISSING.
EXPI Resist.
INSPI Resist.
118/174
Icon
Alarm priority
High priority.
HP
HP
High priority.
V V
High priority.
V V
Icon
Alarm priority
Low priority.
EXT.
Low priority.
Low priority.
Low priority.
Cause
Remedy
Pressure in circuit
Check alarm
higher than High
threshold is
Pressure alarm
correctly set and
threshold.
patient's airway is
clear, and check
patient circuit
integrity.
Tidal volume
Check alarm
higher than upper
threshold is
Tidal Volume
correctly set and
alarm threshold.
check patient
circuit integrity
and expiratory
valve function.
Tidal volume
Check alarm
lower than lower
threshold is
Tidal Volume
correctly set and
alarm threshold.
check patient
circuit integrity
and expiratory
valve function.
Cause
Remedy
External battery
Reinstall external
pack
battery pack or
disconnected.
confirm
disconnection by
pressing Alarm
Silence button.
Loss of external
Check external
DC supply.
DC supply
connection or
confirm
disconnection by
pressing Alarm
Silence button.
Abnormal
Change patient
expiratory
circuit and repeat
resistance in
manual test.
patient circuit.
Abnormal
Change patient
inspiratory
circuit and repeat
resistance in
manual test.
patient circuit.
NOT013189-6