9.3 Technical specifications for accessories

Caution
!
All accessories used must comply with CE marking and meet the specifications
recommended by the manufacturer.
Patient circuit with water traps
Maximal internal volume of circuit: 1000mL.
Compliance: 1.0mL/hPa (= 1.016mL/cmH
Inspiratory resistance at 30L/min: 0.5hPa (= 0.508cmH
Inspiratory resistance at 60L/min: 1.3hPa (= 1.321cmH
Expiratory resistance at 30L/min: 0.5hPa (= 0.508cmH
Expiratory resistance at 60L/min: 1.3hPa (= 1.321cmH
Diameter of the connector: 22mm.
Patient circuit with humidifier
Maximum internal volume of circuit: 1500mL.
Average compliance of circuit: 1.4mL/hPa (= 1.422mL/cmH
Antibacterial filter
Filter: antibacterial / antiviral.
Connector: 22 female and 22 male / 15 female.
Bacterial and viral retention > 99.999%.
Resistance at 30L/min: 1.1hPa (= 1.118cmH
Resistance at 60L/min: 2.2hPa (= 2.235cmH
Compliance: 0.2mL/hPa (= 0.203mL/cmH
Compressible volume: 66mL.
Internal volume: 200mL.
Usage time: 24 hours.
Humidifier
CE-marked equipment. The humidifier used must meet the following requirements:
Maximum pressure during operation > 80hPa.
Max. flow: 180L/min.
Pressure drop at 180L/min: < 3hPa.
During spontaneous breathing: < 3hPa.
Air loss at maximum pressure: < 20mL/min.
Average compliance: 0.3 - 0.5mL/hPa.
Liquid flow rate: 10 - 25mg/L.
Anti-dust filter
Specific ResMed equipment, ref. MOU010645.
Note: The manufacturer accepts no liability for damage to the device caused by the use of a
filter which does not comply with the recommendations.
Remote alarm
Specific ResMed equipment, ref. BOI010525.
Note: The manufacturer accepts no liability for damage to the device caused by connection to a
remote alarm which does not comply with the specifications provided.
146/174
O).
2
O).
2
O).
2
O).
2
O).
2
O).
2
O).
2
O).
2
O).
2
NOT013189-6