7.3 General alarms causes and remedies

Table 7.3: Audible alarms which cannot be turned off

Message
INT BATT EMPTY.
MINI FiO2.
MAXI FiO2.
NiMH
BATTERY.
Table 7.4: Alarms which can be turned off temporarily (for 120 seconds after Alarm Silence button
pressed)
Message
REPLACE
O2 CELL.
PATIENT
APNEA.
EXT BATT
EMPTY.
EXT BATT
LOW.
116/174
Icon
Alarm priority
High priority.
INT.
High priority.
O O
2
High priority.
O O
2
Low priority.
Icon
Alarm priority
Low priority.
Medium priority.
Low priority.
EXT.
Low priority.
EXT.
Cause
Remedy
Internal battery
Change power
charge less than
supply.
5%.
Supplied FiO
too
Adjust rotameter
2
low.
value or Minimum
FiO
2
threshold.
Supplied FiO
too
Adjust rotameter
2
high.
value or
Maximum FiO
alarm threshold.
Connection of
Connect
unknown external
compatible
battery pack.
external battery
pack (Li-Ion-type).
Cause
Remedy
O
Fi
cell fault.
Change external
2
sensor and
perform manual
test to measure
new sensor
characteristics, or
turn off O
No spontaneous
Adjust inspiratory
cycles initiated
trigger value.
during Tapnea
starting from end
of previous
expiratory phase.
External battery
Connect device to
charge less than
mains supply or
5%.
change external
battery pack.
External battery
Connect device to
charge less than
mains supply or
15%.
change external
battery pack.
alarm
2
option.
2
NOT013189-6