Pressure
Maximum pressure:
60hPa (= 60.96cmH
60hPa (= 60.96cmH
To obtain these maximum pressures:
For pressure ventilation:
Set up PACV with Pinsp = 50cmH
patient circuit fitted with a Maquet lung test to the device. Start ventilation.
For volumetric ventilation:
Set up adult, non-invasive ACV and set V
60.96cmH
ventilation and check that the HP alarm is triggered.
Maximum pressure in single-fault conditions: 70hPa (= 71.12cmH
Negative pressure during exhalation: None.
The High Pressure alarm is triggered when the inspiratory proximal pressure is higher than
the High Pressure alarm threshold.
As a safety feature, if the turbine output pressure is greater than 60 – 70hPa (= 61.0 –
71.1cmH
O), the 30V power supply to the turbine is cut. This is achieved by means of an
2
electronic pressure switch which is independent of both the microprocessor and the proximal
pressure sensor.
Resistance
Ventilator with double patient circuit (Ø22 mm) ref. CIR009727
Inspiratory resistance at 60L/min: 0.7 hPa (= 0.711cmH
Expiratory resistance at 60L/min: 1.7 hPa (= 1.727cmH
Inspiratory resistance at 30L/min: 0.4 hPa (= 0.406cmH
Expiratory resistance at 30L/min: 1.0 hPa (= 1.016cmH
Ventilator with double patient circuit (Ø22 mm) ref. CIR009727 & accessories
(tubing, filter, humidifier)
Device volume: 1,700mL.
Device compliance: 1.7mL/hPa (= 1.727mL/cmH
Inspiratory resistance at 60L/min: 4.3hPa (= 4.369cmH
Expiratory resistance at 60L/min: 3.0hPa (= 3.048cmH
NOT013189-6
O) in volumetric ventilation;
2
O) in pressure ventilation.
2
O). Connect a double limb patient circuit and block the other end. Start
2
O and PEEP = 10cmH
2
2
= 0.65L and Pmax alarm = 60hPa (=
T
O).
2
O).
2
O).
2
O).
2
O).
2
O).
2
O).
2
Technical Specifications and Description
O, and connect a double limb
O).
2
143/174