Figure 5.8: Stage 4 of the manual test.
The ventilator detects the circuit type and the resistance and compliance of the patient circuit,
and calibrates the expiratory spirometry.
Figure 5.9: Results of the manual test.
The results of the test are displayed. The results show whether or not the test was successful,
the type of circuit connected and the resistance and compliance values of the patient circuit.
68/174
Obstruct the end of the
patient circuit
When done, press :
CONTINUE
CONTINUE
Last test
the 12/06/07 at 14 : 50 : 35
Circuit test : PASS
Kind of circuit : DOUBLE
Ctot = 1.2 ml/cmH2O
Ri = 0.2 cmH2O (30 l/min)
Ri = 0.2 cmH2O (60 l/min)
Re = 0.2 cmH2O (30 l/min)
Re = 0.2 cmH2O (60 l/min)
RESTART
RESTART
Press to cancel test
Cancel
Press to continue test
Press to return to the main
Back
screen
Press to restart test